10
§ Installing the connecting cable:
– Remove the plug from the bore hole on the indicator.
If the same cable is used for COM1 connections (terminal screw LV1), use the bore hole
in the middle of the housing.
– Guide the cable with the attached cable gland through the bore hole.
– Close and tighten the cable gland in accordance with the applicable standards.
– Make sure the shielding has contact with the clamps, as the shielding provides the
grounding connection.
§ Connecting the wires inside the indicator:
– Connect the wires securely in accordance with the terminal assignments.
§ Close the indicator. Make sure that the rubber seal between the front panel and the
housing is correctly positioned.
§ After you close the housing again, use a pressure gauge to check the integrity of the
IP67-protection. For details, contact the Sartorius Service Center.
Configuring the UniCOM Interface for Operation
After configuring the desired operating mode and installing the interface module in the
indicator, configure the interface by selecting the settings in the operating menu that
correspond to your requirements. For details on the settings available in the operating
menu, see "Configuring the UniCOM Interface" below. For additional information, see
also the chapter entitled "Settings" in the Combics operating instructions.