Display Symbols; Error Messages And Troubleshooting - Brecknell MS140/300 Mode D'emploi

Balance électronique médicale portable à rampe
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7.2 Display Symbols

Symbol
0------
0_______
Ad------
Ad______
--------
________
eep. e 0
EEP. E 1
EEP. E 2
Lo. b At
CAP. - -
CAL. P x
CAL. E r
d. - -
px. y

7.3 Error Messages and Troubleshooting

Symptom
Ad------
Loadcell wires to indicator are incorrectly
connected, shorted, opened, ADC or loadcell(s)
Ad______
are damaged.
0------
Weight reading exceeds Power On Zero limit.
0______
Weight reading below Power On Zero limit.
--------
Weight reading exceeds Overload limit or the
weight value cannot be displayed in the current
unit of measure because it exceeds 6 digits.
________
Weight reading below under load limit.
EEP. E 2
Parameter is not correctly set.
CAL. E r
Calibration error. Input data or loaded weight is
too small, too big, unstable, non-linear.
Will not turn on.
Power cord not plugged in or properly
connected. Power outlet not supplying
electricity. Battery discharged. Other failure.
Cannot zero the display
Load on scale exceeds allowable limits. Load
or will not zero when
on scale is not stable. Load cell damage.
turned on.
33
Zero is over the setting range
Zero point is below the setting range
Signal to ADC is over maximum range
Signal to ADC is below minimum range
Weight is over upper limitation or display data is over limitation
Weight is below lower limitation
EEPROM can't be accessed.
The parameters are not the same as the backup data
The setting parameter(s) is not in normal range
Battery voltage is below 4.7V
The setting full capacity will be displayed
Calibration on point (x)
Calibration error, maybe input data or loaded weight is incorrect, unstable, non-linear
The division will be displayed
The x = parameter, y = parameter choice
Probable Cause
Description
Remedy
Make sure wires are ok and correctly connected.
Replace loadcell or ADC chip, Service required.
Make sure scale platform is empty. Perform zero
calibration.
Install platform on scale. Perform zero calibration.
Reduce load on scale until weight value can be
displayed. Use a more appropriate unit of measure.
Install platform on scale. Perform zero calibration.
Re-set parameter
Input correct data, load correct weight onto
platform, Service required.
Check power cord connections. Make sure power
cord is plugged into the power outlet. Check power
source. Replace batteries. Service required.
Remove load on scale. Wait for load to become
stable. Service required.