Brecknell MS140/300 Mode D'emploi page 31

Balance électronique médicale portable à rampe
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MS140/300 User Instructions
Table 6.2 Symbols Used
<LF>
Line Feed character (hex 0AH)
<CR>
Carriage Return character (hex 0DH)
<ETX>
End of Text character (hex 03H)
<SP>
Space (hex 20H)
H
H
H
Four current status bytes
1
2
3
<P>
Polarity character: "?" or " "
W
---W
Reading data, 1-7 bytes (seven digits)
1
7
<DP>
Decimal point
U
U
Measure units, kg, lb, lb:oz, % or pcs; 2-5 bytes
1
2
Table 6.3 Bit Definition of H
Byte 1 (H
)
Bit
1
0
0 = stable
0 = not under capacity 00 = compare disable
1 = not stable
1 = under capacity
1
0 = not at zero point 0 = not over capacity
1 = at zero point
1 = over capacity
2
always 0
always 0
3
0 = eeprom OK
always 0
1 = eeprom error
4
always 1
always 1
5
always 1
always 1
6
always 0
always 1
7
parity
parity
H
H
1
2
3
Byte 2 (H
)
Byte 3 (H
2
01 = lower limit
10 = ok
11= upper limit
0 = gross weight
1 = net weight
always 0
always 1
always 1
always 0
parity
)
3
29

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Ps-500

Table des Matières