SubMenu1
Option
P5
0-4
p6
0-2
p7
00-31
P8
0-2
p9
0-5
p10
0,1
p11
0-6
p12
0-7
p13
0-7
p14
0-2
MS140/300 User Instructions
Default
Parameter Description
3
Baud rate for RS-232
0
RS-232 protocol
9
Scale resolution select
1
Calibration division size
1
Decimal point in calibration Refer to
1
Calibration unit
6
Weigh units enable
3
Power on zero range
2
Zero range for [ZERO] key 0 = power-on zero-point +1%FS
2
Weight signal within power
on zero point range
Comment
0 = 1200 bps
1 = 2400 bps
2 = 4800 bps
3 = 9600 bps
4 = 19200 bps
0 = 8NO
1 = 7O1
2 = 7E1
Refer to
Table 4.3
for resolution choices and
values.
Refer to
Table 4.4
and
Table 4.5
0 = 1
1 = 2
2 = 5
Table 4.4
and
Table 4.5
0 = 1
1 = 0.1
2 = 0.01
3 = 0.001
4 = 0.0001
5 = 10
Refer to
Table 4.4
and
Table 4.5
0 = kg
1 = lb
0 = only kg
1 = only lb
2 = only lb:oz
3 = kg or lb
4 = kg or lb:oz
5 = lb or lb:oz
6 = kg, lb, or lb:oz
0 = calibration zero -point +1%FS
1 = calibration zero -point +2%FS
2 = calibration zero-point +5%FS
3 = calibration zero-point +10%FS
4 = calibration zero-point +20%FS
5 = calibration zero-point +50%FS
6 = calibration zero-point +100%FS
7 = no limitation
1 = power-on zero-point +2FS
2 = power-on zero-point +5%FS
3 = power-on zero-point +10%FS
4 = power-on zero-point +20%FS
5 = power-on zero-point +50%FS
6 = power-on zero-point +100%FS
7 = no limitation
0 = current weight
1 = calibration zero-point
2 = switch off zero point and power off tare
weight as current tare weight
17