TECHNICAL DATA
Type
Measurement method
Measurement range
Measurement accuracy
Wrist circumference
Power supply
Number of measurements
Classification
Clinical test
Wireless communication
Operating conditions
Transport/Storage conditions
Dimensions
Weight
Ingress protection
Applied part
Useful life
E-30
UB-1100CNBLE
Oscillometric measurement
Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse: 40 - 180 beats/min
Pressure: ±3 mmHg
Pulse: ±5 %
13.5cm - 21.5cm
3.7V Li-ion 325mAh
Approx. 100 times
With pressure value 170 mmHg,
room temperature 23 °C.
Internally powered ME equipment
(Supplied by batteries)/
Class II (Supplied by adapter)
Continuous Operation mode
According to ISO81060-2
ZY (MURATA Manufacturing Co.Ltd.)
Bluetooth
Ver. 4.1 Low Energy
®
Technology BLP
+10 to +40 °C / 15 to 85% RH / 800
to 1060 hPa
-20 to +60 °C / 10 to 95% RH / 700
to 1060 hPa
Approx. 54 [W] x 76 [H] x 21 [D] mm
Approx. 92g
Device: IP20
Cuff Type BF
5 years (when used six times a day)