Technical Data - A&D Medical Life Source UA-767FAM Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Type
Measurement method
Measurement range
Measurement accuracy
Power supply
Number of
measurements
Classification
Clinical test
EMC
Memory
Operating conditions
Transport/Storage
condition
Dimensions
Weight
Applied part
Useful life
Ingress protection

Technical Data

UA-767FAM
Oscillometric measurement
Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse:
40 - 180 beats / minute
Pressure: ±3 mmHg
Pulse:
±5%
4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233)
UA-767FAM includes an AC adapter.
Approx. 700 times LR6 (alkaline batteries)
Approx. 200 times R6P (manganese batteries)
With pressure value 180 mmHg, room temperature
73.4 °F (23 °C)
Internally powered ME equipment (Supplied by
batteries) / Class II (Supplied by adapter)
Continuous operation mode
According to ANSI / AAMI SP-10 1992
IEC 60601-1-2: 2007
Last 60 measurements for each of the four user
+50°F to +104°F (+10°C to +40°C) / 15 to 85 %RH /
800 to 1060 hPa
-4°F to +140°F (-20°C to +60°C) / 10%RH to
95%RH
Approx. 5.5" (140mm) [W] x 2.3" (60mm) [H]
x 4.1" (105mm) [D]
Approx. 9.0 oz. (255 g), excluding the batteries
Cuff:
Type BF
Device:
5 years (when used six times a day)
Cuff:
2 years (when used six times a day)
Device: IP20
English 23