Table A-2. Electromagnetic Immunity - General
IEC 60601
Immunity Test
Test Level
Electrostatic
+ 6 kV contact
discharge (ESD)
+ 8 kV air
IEC 61000-4-2
Electrical fast
+ 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
+1 kV for input/
output lines
Surge
+ 1 kV differential
mode
IEC 61000-4-5
+ 2 kV common
mode
Voltage dips, short
< 5% U
interruptions, and
(> 95% dip in U
voltage variations
for 0,5 cycle
on power supply
input lines
40% U
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
< 5% U
(> 95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
Magnetic field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
Compliance
Level
+ 6 kV contact
+ 8 kV air
+ 2 kV for power
supply lines
+1 kV for input/
output lines
+ 1 kV differential
mode
+ 2 kV common
mode
< 5% U
T
T
)
(> 95% dip in U
T
for 0,5 cycle
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
< 5% U
T
T
)
(> 95% dip in U
T
for 5 sec
3 A/m
Electromagnetic
Environment - Guidance
Floors should be wood,
concrete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
)
T
hospital environment. If the user
of the monitor requires continued
operation during power mains
interruptions, it is recommended
)
T
that the monitor be powered from
an uninterruptible power supply
or a battery.
)
T
)
T
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Electromagnetic Compatibility
A-3