Error Messages; Safety, Care, Accuracy Test And Disposal; Safety And Protection; Device Care - Microlife Automatic BP A2 Easy Mode D'emploi

Masquer les pouces Voir aussi pour Automatic BP A2 Easy:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.

9. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«ERR 2»
Error signal During the measurement, error signals
were detected by the cuff, caused for
AO
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«ERR 3»
No pres-
An adequate pressure cannot be gener-
sure in the
ated in the cuff. A leak may have
AN
cuff
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
300 mmHg) OR the pulse is too high
high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
BP A2 Easy

10. Safety, Care, Accuracy Test and Disposal

Safety and protection

 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
 The cuffs are sensitive and must be handled with care.
 Do not exchange or use any other kind of cuff or cuff connector
for measuring with this device.
 Only pump up the cuff once fitted.
 Do not use this device close to strong electromagnetic fields
such as mobile telephones or radio installations.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 If the device is not going to be used for a prolonged period the
batteries should be removed.
 Read the additional safety information in the individual sections
of this instructions.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!

Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact
Microlife-Service to arrange the test (see foreword).
EN
5

Publicité

Table des Matières
loading

Table des Matières