Bühler technologies GAS 222.21 Notice De Montage page 13

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
GAS 222.21
5.2.1 Menu Navigation Overview
Temperature and status display
Display
____
The probe temperature is indicated in 0.5°C increments. Press the Enter key to access the main menu.
The temperature can be displayed in Celsius or Fahrenheit. During operation the conditions of the
current temperature
blowback controller (optional) is showed alternating with the probe temperature: Blowback active and manual operation
(refer to the description of the blowback controller or the relevant menu operation).
Main menu
Display
Probe controller
*
* These menus
will only appear
if one of the
options is ins
talled.
*
Top Settings
Global settings
Exit
Exit main menu
BX460046 ◦ 09/2021
Submenu
Prob
Temperature
Target temperature
Alarm
Alarm hysteresis
Exit
Exit submenu
* Regulator for
Temperature
heated
Target temperature
extension
(add-on)
(optional)
Display:
AdOn
Alarm
Alarm hysteresis
Exit
Exit submenu
* Blowback
Control mode
controller
Control mode
(optional)
Display:
bbc
Time of interval
Blowback interval
Time of blowback
Blowback time
Exit
Exit submenu
toP
E
Setting range
tEMP
Selection menu
Probe temperature
ALAr
Selection menu
Setting range
E
tEMP
Selection menu
Extension temperature
ALAr
Selection menu
Setting range
E
ctrl
Selection menu
Auto - Hand
tInt
Selection menu
5 min...99h59min
t_bb
Selection menu
5...59 s
E
Selection menu
°C - °F
Adjust value
stores the value
wait 5s: do not save
Setting the probe temperature
____
within the range:
50 °C...200 °C / 122 °F...392 °F
___
Setting the alarm hysteresis
within the range: ±5°C...±30°C
Setting the temperature of the
____
heated extension within the range:
50 °C...200 °C / 122 °F...392 °F
___
Setting the alarm hysteresis
within the range: ±5°C...±30°C
Auto
Switching from automatic
to manual blowback
__:__
Setting the blowback
interval time
__. 0
Setting the blowback time
within the range 5 s to 99 s
Setting the display unit.
C
Available in Celsius or
Fahrenheit
Bühler Technologies GmbH
13