WARNING
• Immediately after a surgery, the handpiece should be cleaned, lubricated and sterilized. Failure to properly maintain the
handpiece can lead to contaminants remaining inside the handpiece. Unremoved blood may result in product failure due to
the internal blood coagulation. Improper maintenance may also lead to overheating, causing burn injuries, product failure, or
difficulty in attaching and detaching the bur. (Refer to "8. Post-use Maintenance")
• Depressing the Push Button while the handpiece is in rotation may lead to overheating, causing burn injuries or product
failure. Avoid the push button to contact with any oral tissue.
• Supply coolant water through the spray nozzle while using the handpiece. Other water supply methods than recommended
may lead to overheating causing burn injuries, or product failure.
• Do not use the handpiece for prophylaxis purposes. Polishing paste could enter the handpiece, causing the failure of the
Push Button or burn injuries due to the overheating of the handpiece head.
• Keep any debris or other foreign materials away from inside the gear or the handpiece. Foreign materials remaining inside
may lead to overheating, causing burn injuries or other accidents.