Chuck cleaning
Clean the Push Button chuck once a week.
1) Mount the Tip Nozzle into the spray can port.
2) Lubricate the chuck directly through the bur insertion hole (Fig. 9).
3) Lubricate handpiece by using NSK PANA SPRAY Plus (Fig. 7) or NSK automatic handpiece cleaning and
lubrication system.
CAUTION
• If the chuck is not regularly cleaned the chuck grip may be weakened and the bur may be accidentally released during use.