3) Insert the Purge Nozzle into the rear of the handpiece until it is locked in place with a clicking sound
(Fig. 8
).
4) Insert an air syringe/air blow gun into the Purge Nozzle. Then blow air for 30 seconds or more to purge
excess oil inside the handpiece (Fig. 8
WARNING
• Excess oil remaining inside the handpiece may cause overheating of the
handpiece which could lead to a burn injury.
CAUTION
• Debris and blood could enter the product during surgical operations. To avoid accumulation of debris inside the product, NSK recommends
lubrication with PANA SPRAY Plus immediately after use and before sterilization.
• When applying spray be sure to hold the handpiece firmly to prevent the handpiece from slipping out of the hand due to the spray pressure.
• Hold the spray can upright.
18
).
2
Air Syringe
1
Purge Nozzle
Fig. 8