7-7 Sterilization
The only recommended sterilization method is by steam autoclaving. Remove the bur after each patient and sterilize as follows.
1) Place the handpiece into an autoclave pouch and seal.
2) Autoclave according to the following time requirements.
Autoclave for more than 20 min. at 121˚C, 15 min. at 132˚C.
3) The handpiece should remain sealed in the autoclave pouch until required for use.
CAUTION
· Do not autoclave the handpiece with other instruments even in a pouch. This is to prevent possible discoloration and damage to the handpiece
from chemical residue on other instruments.
· Keep the handpiece in normal atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
· Immediately after use the handpiece should be cleaned, lubricated and sterilized. If blood remains on any external or internal surfaces it can
harden and cause rust.
· Do not heat or cool the handpiece too quickly. Rapid change in temperature could cause damage to the handpiece.