NSK Ti-Max X450L Manuel D'opération page 13

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

X450M4 / X450 5H
1) Remove the handpiece from the hose.
2) Insert the arrow-head spray nozzle tip into the spray can. (Fig. 11)
3) Place the Spray Nozzle into the Drive Air Port of the handpiece and spray for approximately 2-3
seconds. Continue spraying until it runs clear from the handpiece head for at least 2 seconds. (Fig. 12)
CAUTION
· When applying spray be sure to hold the handpiece firmly to prevent the handpiece from slipping out
of the hand due to the spray pressure.
· Hold the spray can upright.
12
Arrow-head Spray Nozzle
Handpiece
PANA SPRAY Plus
Fig. 11
Midwest 4-hole
5-hole Joint
Drive Air Port
Drive Air Port
Fig. 12

Publicité

Table des Matières
loading

Table des Matières