Discontinuing A Measurement; Delete Memory; Error Messages/Malfunctions - Marque Verte Digitensio PRO Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Further information
Measurements should not occur soon after each other, since otherwise, the results will
be falsified. Wait therefore for several minutes in a relaxed position, sitting or lying,
before you repeat a measurement.

5.6. Discontinuing a measurement

If it is necessary to interrupt a blood pressure measurement for any reason
(e.g the patient feels unwell), the
device then immediately lowers the cuff pressure automatically.

5.7. Delete memory

To delete stored memory readings, hold on «MEMORY»
button down till LCD shows a symbol of «CL» flashing. Then
press memory button again, you will hear 3 short beeps
indicating memories have been deleted.

6. Error messages/malfunctions

If an error occurs during a measurement, the measurement is dis-
continued and a corresponding error code is displayed
(Example: Error No. 2).
Error No. Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason:
The arm was moved during the measurement (Artefact).
ERR 3
The inflation of the cuff takes too long. The cuff is not correctly seated.
ERR 5
The measured results have indicated an unacceptable difference between the
systolic and diastolic pressure. Take another measurement, carefully following
the instructions. Contact your doctor, if you still obtain unusual results.
ERR 6
Single data differ too much during Average Mode even after 4 cycles. No
average result can be displayed.
Size: 17mm * 60mm
button can be pressed at any time. The
1
34

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Bp 3bd1-4d

Table des Matières