In case that the single data of each cycle differ too much from each other, a fourth measurement is
performed before the result will be displayed. In rare cases the blood pressure is such unstable that even
after four measurements the data vary too much. In this case «ERR 6» is shown and no result can be given.
If one of measurement causes an error message it is repeated.
5.5. Memory-storage and recall of measurements
The blood-pressure monitor automatically stores the last 30 measurement values for each of 2 users. By
pressing «MEMORY» button, the screen will show the most recent readings along with time and date.
Each time you press the memory button an earlier measurement will be displayed.
If an irregular heartbeat interval was detected in any measurements stored in memory, the
be displayed.
Further information
Measurements should not occur soon after each other, since otherwise, the results will be
falsified. Wait therefore for several minutes in a relaxed position, sitting or lying, before you
repeat a measurement.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any
reason (e.g. the patient feels unwell), the
at any time.
The device immediately lowers the cuff pressure automatically.
5.7. Delete memories
To delete stored memory readings, hold on memory button down till LCD
shows a symbol of «CL» flashing. Then press «Memory» button again, you
will hear 3 short beeps indicating memories have been deleted.
Before you delete all memory readings stored in, make sure you won't need
refer to them at a later date. Keeping a written record is prudent and may
provide additional information for your doctor's visit.
28
indicates measurement was taken under average mode.
1
1
mmHg
mmHg
mmHg
mmHg
Memory reading
button can be pressed
icon will