8.2 Data transmission (models with a WiFi interface)
Problem
... the status LED does not come on?
... only the "SYS" item is visible in the "rF"
menu?
... after calling up the menu, the "rF" item is
not displayed?
... the display noChG appears when the WiFi
interface is activated?
... the status LED is illuminated red?
... the display Id:Err appears?
... Enter is pressed and the display Err:71
appears?
8.3 Changing batteries
Press!
68 •
The status LED is faulty.
- Inform seca Service
The WiFi interface is deactivated.
- Activate WiFi interface
The scale's WiFi interface is faulty.
- Inform seca Service
Scale in battery mode, On option selected in rF\SYS menu
- Select PErM option
- Use power supply unit (recommended)
• ID
found in hospital information system (HIS) or in seca software (see
not
fault
).
Id:Err
• Device has
saved measured results to the clipboard.
not
- Repeat measurement
• Measured results have
- Deactivate BMIF function (see "Using the differential value function
(bMIF)" on page 61)
- Repeat measurement
- Check WiFi connection
The ID of the user or the patient has not been found in the HIS or in the
seca software.
- Scan the barcodes for the user and patient again
- Inform seca Service
No data transmission possible, WiFi interface is deactivated.
- Activate WiFi interface (see "Setting up the WiFi connection (336 i)" on
page 55)
You need 6 AA-type batteries, 1.5 V. To establish the power supply, proceed
as follows:
1. Open the battery compartment.
2. Take the discharged batteries out of the battery compartment.
3. Put the new batteries in the battery compartment.
NOTE
Ensure the correct polarity of the batteries (markings on the battery
holder). If the display
teries the wrong way round or the batteries have discharged. If batter-
ies are inserted the wrong way round, they must be removed again
immediately.
4. Close the battery compartment.
Cause/solution
been sent to an HIS or to seca software.
not
appears, you have inserted one of the bat-
bAtt