Measuring rod (if present)
6.3 Sterilizing
7. FUNCTION CHECK
66 •
3. Disinfect the device:
►
Moisten a soft cloth with disinfectant and wipe down the device.
►
Comply with the intervals: see table.
Interval
every measurement
Before
every measurement
After
If required
1. Check that your disinfectant is suitable for sensitive surfaces and acrylic
sheet (active ingredient: quaternary ammonium compounds, for example).
2. Follow the instructions on the disinfectant.
3. Disinfect the device:
►
Moisten a soft cloth with disinfectant and wipe down the device with
it.
►
Comply with the intervals, see table.
Interval
every
Before
measurement
every
After
measurement
If required
This device may not be sterilised.
►
Perform a function check prior to each use.
A complete function check includes:
• visual inspection for mechanical damage
• checking the alignment of the device
• visual and function check of the display elements
• function check of all the controls shown in the section entitled
"Overview"
• function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled "What do I do if ..." in this
document.
Component
Tray
Tray
Display elements,
controls, housing
Component
Head stop, foot stop
Head stop, foot stop
• Controls
• Housing
• Measuring rod with scale