Endress+Hauser iTEMP HART TMT142 Manuel De Mise En Service page 65

Transmetteur de température
Masquer les pouces Voir aussi pour iTEMP HART TMT142:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TMT142
Sensor type
Sensor type
GOST
Polynomial RTD
Callendar - van
Dusen (Pt100)
TC Type B
TC Type C
TC Type D
TC Type E
TC Type J
TC Type K
TC Type L
TC Type N
TC Type R
TC Type S
TC Type T
TC Type U
10 to 400 Ω
10 to 2000 Ω
-20 to 100 mV
Specific linearisation and sensor matching
Selecting the sensor types 'Callendar-van-Dusen' or 'Polynomial RTD' improves the accuracy of the system or defines user-specific linearisation of resistance
thermometers. A detailed description of the 'Callendar-van-Dusen' method and 'Polynomial RTD' linearisation is provided in the Appendix to these Operating
Instructions.
!
The selection list of the sensor type is displayed depending on the PV unit. Example: When
selecting a resistance thermometer the PV unit must first be set to Ω.
Sensor connection
Input of RTD connection mode.
Input:
• 2-wire
• 3-wire
• 4-wire
!
Function is only active on selection of a resistance thermometers (RTD) in the device
function SENSOR TYPE.
Cold junction
Selection of the internal (Pt100) or an external comparison
measurement point.
Input:
• internal
• external
!
Function is only active on selection of a thermo-couple (TC) in the device function SENSOR
TYPE.
External
Input of the external comparison point measurement value.
temperature
Input: -40.00 to 85.00 °C (°C, °F, K)
0 °C
!
Function is only active when "external" has been selected in the device function COLD
JUNCTION.
Endress+Hauser
Meas. range
start
Pt50
-200 °C (-328 °F)
Pt100
-200 °C (-328 °F)
Cu50
-200 °C (-328 °F)
Cu100
-200 °C (-328 °F)
-200 °C (-328 °F)
-200 °C (-328 °F)
0 °C (32 °F)
0 °C (32 °F)
0 °C (32 °F)
-270 °C (-454 °F)
-210 °C (-346 °F)
-270 °C (-454 °F)
-200 °C (-328 °F)
-270 °C (-454 °F)
-50 °C (-58 °F)
-50 °C (-58 °F)
-270 °C (-454 °F)
-200 °C (-328 °F)
10 Ω
10 Ω
-20 mV
Note!
Note!
Note!
Note!
Function group SENSOR
Meas. range
full scale value
1100 °C (2012 °F)
850 °C (1562 °F)
200 °C (392 °F)
200 °C (392 °F)
850 °C (1562 °F)
850 °C (1562 °F)
1820 °C (3308 °F)
2320 °C (4208 °F)
2495 °C (4523 °F)
1000 °C (1832 °F)
1200 °C (2192 °F)
1372 °C (2501.6 °F)
900 °C (1652 °F)
1300 °C (2372 °F)
1768 °C (3214.4 °F)
1768 °C (3214.4 °F)
400 °C (752 °F)
600 °C (1112 °F)
400 Ω
2000 Ω
100 mV
min. range
+
10 K (18 °F)
10 K (18 °F)
10 K (18 °F)
10 K (18 °F)
10 K (18 °F)
10 K (18 °F)
500 K (900 °F)
500 K (900 °F)
500 K (900 °F)
50 K (90 °F)
50 K (90 °F)
50 K (90 °F)
50 K (90 °F)
50 K (90 °F)
500 K (900 °F)
500 K (900 °F)
50 K (90 °F)
50 K (90 °F)
10 Ω
100 Ω
5 mV
+
+
+
Commissioning
+
+
+
+
65

Publicité

Table des Matières
loading

Table des Matières