Availability in ReadWin
2000, HART
Measuring point
Input and display of the information relating to the measuring point identification
Measuring point
Input: 8 characters
TAG
Description
Input: 16 characters
Message
Input: 32 characters
Device information
Display of the information relating to the device identification
Device release
Display of device release
Serial number
11 digit display of the E+H device serial number
(equal to that on the legend plate).
Software version
Display of the software version
Hardware version
Display of the hardware version
Certificates
Display of device approvals
Device
Display of the information relating to the HART
Manufacturer
Manufacturer's identification character: Endress+Hauser (=17)
Device type
Device type identification: TMT 142
Date
Individual use of this parameters
Hardware revision
Revisions of the devices' electronic components
®
Availability in ReadWin
2000, HART
Security locking
Set-up release code.
Input:
• Lock = 0
• Release = 241
Reset to default
Reset to factory default values.
Input: 142
0
Output simulation
Activate simulation mode.
Input:
• OFF
• ON
Simulation value
Input of the simulation value (current).
Input: 3.58 to 23 mA
Endress+Hauser
Function group IDENTIFICATION
®
handheld terminal DXR 375 (symbol
®
device identification
Function group SERVICE FUNCTIONS
®
handheld terminal DXR 375 (symbol
7
)
7
)
Commissioning
7
®
ReadWin
2000
/FieldCare
+
+
+
+
-
+
-
+
+
+
+
+
+
+
-
+
-
+
-
+
-
+
-
+
7
®
ReadWin
2000
/FieldCare
+
+
+
+
+
+
+
+
69

Publicité

Table des Matières
loading

Table des Matières