ERROR MESSAGES
During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
ERROR
DESCRIPTION
CODE
Time input Error
Err 1
Date input Error
Err 2
Initial Zero is greater than
Err 4
allowed (4% of maximum
capacity) when power is
turned on or when the
[Zero/Enter] key is pressed.
A/D count is not correct
Err 6
when turning the scale on.
Percent input error
Err 7
High limit input error
Err 8
© Adam Equipment Company 2011
POSSIBLE CAUSES
Invalid time entry such as "268970"
for the time format "H-m-S".
34th day of a month is an invalid
entry.
Weight on the platform when turning
the scale on.
Excessive weight on the platform
when zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Load cell is damaged.
Electronics is damaged.
Percent function is entered with no
reference mass on the platform.
Low limit is set first, then the high
limit is set lower than the low limit
and high limit not equal to zero.