4 Using Cassettes and Preparing Instruments
4.1 Using the STATIM 2000 G4 Cassette
When removing the cassette after a cycle, exercise caution as the metal areas will be hot and the
cassette may contain hot steam.
Opening the Cassette:
1. Hold the cassette handle with your thumbs facing inward on
the cassette latch.
2. Push downward on the cassette latch.
3. Raise the cassette lid upwards and disengage the hinge.
4. Rest the lid on its outer surface.
Closing the Cassette:
1. Align the hinge tab on the cassette lid with the hinge slot on the rear of the bottom tray.
2. As you begin to close the lid, the hinge tab and slot will engage.
•
Inserting the Cassette into the STATIM 2000 G4:
1. Place the end of the cassette into the unit.
2. Gently push inward until you hear a "click" sound and check the LCD icon for change
from
to
.
Never force the cassette into the STATIM as the interior components could be damaged.
NOTE: The main menu screen will display
Removing the Cassette:
1. Grasp the handle with two hands and pull away from the unit.
2. Pull the cassette clear of the unit and set down on a firm surface.
•
Disengaging the Cassette:
When not in use, the cassette should be disengaged. To disengage the cassette, grasp the handle
and pull the cassette out until there is a 15 mm to 20 mm (½ to ¾") gap between the front of the
STATIM 2000 G4 and the cassette handle.
•
Using the STAT-DRI PLUS Drying Agent
Treatment of the interior surfaces of the cassette with the STAT-DRI PLUS drying agent, provided
with your unit, will enhance the drying process. (Replacement bottles are available from SciCan,
order number 2OZPLUS, 8OZPLUST, 32OZPLUS).
Figure 7
if the cassette is not properly inserted in the unit.
Page 20
cassette latch