SciCan STATIM 2000 G4 Manuel De L'utilisateur page 46

Masquer les pouces Voir aussi pour STATIM 2000 G4:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9 Maintaining your STATIM
Cassette
Biological Filter
Solenoid Valve
Pump
Check Valve
Water Reservoir
Calibration
Technician
Check the tray, lid and seal for damage. Replace if necessary.
Inspect the biological filter for moisture.
Inspect the valve and clean if dirty. Replace the plunger if defective.
Clean the filters, replace if dirty.
Remove the exhaust tube from the back of the unit during the air drying
phase. Check for air coming from the fitting.
Remove the air compressor tube from the check valve inlet while running
a cycle. Make sure no steam is leaking from the valve. Replace if there are
any leaks.
Check the reservoir for dirt. Clean and rinse with steam process distilled
water if necessary.
Calibrate the unit.
Page 46

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Statim 5000 g4

Table des Matières