5 Using your STATIM
5.3.4. Heavy Duty Unwrapped Cycle
(STATIM 5000 G4)
The Heavy Duty Cycle is used to sterilize larger
loads of unwrapped metal instruments weighing
up to 1.5 kg (3.3 lbs). Dental hand pieces can be
sterilized in this cycle. Using the STATIM 5000
Extended Cassette (order no. 01-112509s) will
allow for the sterilization of rigid endoscopes
in this cycle.
HEAVY DUTY UNWRAPPED
HEAVY DUTY UNWRAPPED
Press START
Press START
To select the Heavy Duty Unwrapped Cycle, press the Heavy Duty Unwrapped Cycle button, then press
the START button.
The sterilization temperature in the cassette is 134˚ C (273˚ F) and the holding time is 6 minutes.
HEAVY DUTY UNWRAPPED
Press START
Page 31
D
A
B
C
A
D
Warming up
B
E
Conditioning
F
C
Presurizing
E
F
Sterilizing
Venting
Air Drying