Connection
4
Connection
10
S Route the connection cable of the load cell such that
any water condensate that might occur, or any humid
ity, can drip away downwards. It must not ingress into
the load cell in any way. In the connection area of the
cable end, users must ensure that no humidity can
ingress into the open cable.
S The cable must be routed such that damage to the
external sheath by chafing and/or own movements is
prevented in daily operation.
S The load cells have been designed in four wire
technology; a shortening of the cable is not permissi
ble.
S In order to maintain full precision, cable extensions
should be carried out in six wire technology.
S The cable is suitable for fixed routing, it may not be
bent. The smallest laying radius should not be less
than 300 mm. Tensile straining and twisting of the
cable during and after fitting are not permissible.
Notice
The cable union on the load cell must not be opened un
der any circumstances; if this should occur by mistake,
the load cell must be returned to the manufacturer's
works for repairs.
If elastomer bearings are used, the construction is insu
lated against the foundations. Note in this connection,
DIN/VDE 0100, Part 410, with regard to a necessary po
tential compensation.
A1153-2.0 HBM: public
RTN