Applying the Finger Sensor
M1191T Adult Finger Sensor
For patients weighing more than 50kg.
M1192T Pediatric Finger Sensor
For patients weighing between 15kg and
50kg.
Step
Applying the M1191T (Adult) or M1192T (Pediatric) Sensor
1
Select appropriate sensor for patient size (defined above).
2
Position sensor over patient's finger, being sure the sensor cable
is positioned ON TOP of the finger/hand as shown above.
3
Patient fingertip should touch but not protrude from end of sensor.
If necessary, trim the fingernail to correctly position the sensor.
4
If applying a M1191T Adult finger sensor, secure the cable to
BACK OF HAND with supplied M1627A wristband (wristband is
supplied only with M1191T Adult finger sensors.)
5
Plug sensor into instrument (or into adapter cable if needed).
6
Inspect and change the sensor application site periodically. You
may reapply the sensor to the same patient by repeating this pro-
cedure. Clean and disinfect the sensor before applying it to a dif-
ferent patient.
adult application shown
with supplied wristband