19.3 Adjustment of "verified weighing systems / menu block

Prepare the required adjustment weight. The weight to be used depends
on the capacity of the scale. Carry out adjustment as near as possible to
the scale's maximum weight. Info about test weights can be found on
the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
Call up the navigation in the menu/ function:
 Call up function
Press TARE key
Using navigation keys
select function, (CAL
00  CAL 03)
Back to menu:
Select function "CAL
00" and confirm with
the TARE key.
Overview:
CAL
00
Exit menu
CAL 01
Adjustment with external adjustment weight
CAL 02
Not documented (adjustment with gravitational constant)
KXS/KXG-TM-BA_IA-e-1621