Etac molift Partner 255 Manuel D'utilisation page 11

Masquer les pouces Voir aussi pour molift Partner 255:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 88
Transport and Operating Conditions
Transport and Storage
For long time storage it is recommended that the
emergency stop button is activated (pushed in).
The lifter can be stored and transported under
temperatures between -25 - 70 °C.
Operating
The lifter is designed for use at standard room
temperatures (+5 to +40°C).
Air Pressure: 70 - 106 kPa
Relative Humidity: 15 - 93 %
Following storage or transport at other tempera-
tures leave the lifter in a room with a suitable
temperature until it reaches a safe operating
temperature.
Medical electrical equipment
requires special precautions regard-
ing electromagnetic compatibility
(EMC). Portable or mobile radio
communication equipment may
affect the medical electrical equip-
ment, and should be kept minimum
25 cm (10 inches) from the lifters
electronics.
Batteries
Molift Partner 255 has a 26.4 V 2.6Ah NiMH bat-
tery. The battery has a life expectancy of approx.
500 charge cycles.
Recharge battery upon receipt, and
then every six months, to avoid
battery failure.
Used batteries must be disposed
of as special waste according to
local rules and regulations. Do not
dispose in general waste
Batteries must be fully charged and
depleted completely a couple of
times before it achieves full capac-
x2
ity. Also follow this procedure if
the lifter has not been in use for a
long period (3-4 weeks)
Electronics
Battery and Service indicator
The lifter has two indicator lights on battery
holder.
The electrical system has a power save function
which will turn off the electrical system after ten
minutes without activity. All lights will turn off.
The system is activated when pushing one of the
operating buttons
Battery Light
The battery indicator light will illuminate and
make a sound when battery level is low and the
lifter requires charging. When this occurs the lifter
will have sufficient power available for one full lift-
ing cycle with max load. When battery is critically
low it is only possible to lower the lifting arm.
Service Light
The lifters electronics record the loads lifted and
number of lifts. After a certain period of operation
a signal is given to indicate that service is required.
Service light
No light
Green
Yellow
Red
Red + sound
Service scope
Service involves replacing the lifting motor and
inspection/replacement of worn parts. This must
be carried out by authorized personnel.
Service is needed when the service light is red
(calculated 10.000 lifts in SWL), or service is
required after 5 years if the service counter not has
displayed red yet.
11
Molift Partner 255 / www.etac.com
Wrench symbol /
Service indicator
Battery symbol /
Battery indicator
Mode
Power saving (Stand by)
Ready for use
Order service
Perform service
Perform service
immediately

Publicité

Table des Matières
loading

Table des Matières