Operation
Except where otherwise stated, this specification
applies at an ambient temperature of 20 ºC.
General
Ranges:10 m, 30 m, 100 m, 300 m, 1000 m,
3000m (30 ft, 100 ft, 300 ft, 1000 ft, 3000 ft,
10000 ft)
Accuracy: 1% of range pixel at 0.67 VF
[Note- The measurement accuracy is for the
indicated cursor position only and is conditional
on the velocity factor being correct.]
Resolution:
1% of range.
Input Protection:
This instrument complies
with IEC61010-1 for connection to live systems up
to 300V CAT III. Fused leads must be used if the
voltage between the terminals exceeds 300V.
Output pulse: 5 volts peak to peak into open
circuit. Pulse widths determined by range and
cable impedance:
25Ω
50Ω
10m
7ns
7ns
30m
20ns
20ns
100m
100ns
60ns
300m
300ns
120ns
1000m
1000ns
520ns
3000m
3000ns
2020ns
Gain:
Set for each range with four
user
selectable steps.
Velocity Factor: Variable from 0.20 to 0.99 in
steps of 0.01
Impedance matching: 25, 50, 75, 100
Tx Null (where fitted): 0 to 120
Refresh Rate: Selectable from the menu.Once
per second or three times a
second.
Power Down:
Automatic after 5 minutes with
no key press.
Backlight:
Stays on for 1 minute when
activated.
75Ω
100Ω
7ns
7ns
20ns
20ns
100ns
100ns
170ns
300ns
680ns
1000ns
2340ns
3000ns
15