6.2 Preparation For Operations
Operate the PDE according to the following procedures:
1. Make sure that the power switch has been turned off, and then connect each cable according to Figure 5-1.
2. Put the Camera Unit into the sterile drape when the PDE is being operated in a sterile area or when it is
being used with a patient who has an infectious disease that may be transferred to the Camera Unit. Fix the
sterile drape to the Camera Unit with an attached rubber band.
STERILE DRAPE
Use only the optional sterile drape for the PDE.
Use the sterile drape properly after carefully reading the following descriptions.
6
5
Figure 6-1: Sterile drape
1
Rubber rings
2
Label ② "pull here to stretch drape"
3
Label ① "insert Camera from here"
Guarantee: Three years after the date of sterilization under the conditions of storage in a cool, dry and
dark place.
CAUTION
Instruction Manual PDE
Version 1.30E
1
4
For single use only.
Sterility is guaranteed if the package is unopened and undamaged.
Do not scratch the sterile drape with needles or instruments with sharp edges.
Do not make new holes in the sterile drape.
Store in dark, dry and clean conditions.
The sterile drape is not waterproof. Soaking it in liquid may damage the Camera
Unit.
Do not allow the sterile drape to come into contact with patients.
Make sure that there is no dirt on the camera window before inserting it into the
sterile drape.
Stretched
4
Camera unit
5
Camera hood
6
Cover plate
Operations
2
3
Folded
Page 6-17