Weight
Inspiratory port / double limb adapter
Pressure measurement
Flow measurement
Power supply
External DC Power Supply
Internal Battery
Housing construction
Environmental conditions
3.2 kg (7.1 lb)
22 mm taper, compatible with ISO 5356-1:2004
Anesthetic & Respiratory Equipment – Conical Connectors
Internally mounted pressure transducers
Internally mounted flow transducers
AC 100–240V, 50–60Hz, 90 W 3.75 A continuous, 120 W / 5A peak
110V/400 Hz
12 - 24V DC 90 W, 7.5 A / 3.75 A
Lithium-Ion battery, 14.4 V, 6.6 Ah, 95 Wh
Operating hours (best case): 8 h with a new battery under normal
conditions (see below).
Test conditions: Adult, P(A)CV mode, P control: 20 cmH
15 bpm, Ti: 1.2 sec.
Note: Time may vary with environmental conditions.
Total lifetime: 3,000 hours of operation on internal battery
Operating hours (worst case) > 4 hour run time under the following
conditions:
Test conditions: Adult, non-vented, PACV mode, Double limb circuit,
Pressure Assist = 30 cmH
O, PEEP = 20 cmH
2
Ti: 1.0 sec, Rise Time = Off, Safety Vt = off, Trig = Off. All other parameters
remain at default settings.
Flame retardant engineering thermoplastic
Operating temperature: 32°F to 104°F (0°C to 40°C
Charging temperature: 41°F to 95°F (5°C to 35°C)
Operating humidity: 5 to 93% non-condensing
Storage and transport temperature: -4°F to 122°F (-20°C to 50°C) for up to
24 hours
Storage and transport temperature: -13°F to 158°F (-25°C to 70°C) for
greater than 24 hours
Note: Storing the Astral device at temperatures exceeding 122°F (50°C) for
extended period of time may accelerate battery aging. This will not affect
the safety of the battery or device. Refer to Using the internal battery (see
page 41)
Storage and transport humidity: 5 to 93% non-condensing
It takes 40 minutes* for the device to be ready for use on a patient when
removed from storage at the minimum long term temperature and at an
ambient temperature of 68°F (20°C).
*Assumes that the device is connected to an external AC power.
It takes 60 minutes for the device to be ready for use on a patient when
removed from storage at the maximum long term temperature and at an
ambient temperature of 68°F (20°C).
Air pressure: 1100 hPa to 700 hPa
Altitude: 9842 ft (3000 m)
Note: The performance may be limited below 800 hPa or at altitudes above
6562 ft (2000 m)..
Technical specifications
O, PEEP: Off, Rate:
2
O Rate:20 bpm,
2
English
67