8.3 Adjustment log
This function enables automatic log issue after each adjustment. These logs may be
issued by using an optional printer.
Printout example (KERN YKB-01N):
Make sure that the communication parameter of balance and printer are the
same.
Call function
In weighing mode press the CAL key for 3 sec.
Use the navigation buttons ( ) to select „GLP.OUT".
The current setting is marked by the stability display ( )
With stability display ( )
Without stability display ( )
26
- - - - - - - - - - - - - - - - -
CAL –EXTERNAL
- - - - - - - - - - - - - - - - -
KERN & Sohn GmbH
TYPE ABJ 220-4NM
SN WBIIAB000I
ID 1234
REF 200.0000g
BFR 200.0001g
AFT 200.0000g
-COMPLETE
-SIGNATURE-
- - - - - - - - - - - - - - - - -
Function activated
Function disabled
Mode of adjustment
Company
Model
Serial no.
Balance identification no. (see chpt. 8.4)
Used adjustment weight
Before adjustment
After adjustment
prepared by