11. TECHNICAL DATA

Model/Type reference:
Automated BP Specification:
Patient population:
BP Measurement:
Measurement Range:
BP Read Time:
Accuracy:
Heart Rate Accuracy:
Manual SphygMode BP Specification:
Deflate Rate:
Cuff Specifications:
Rechargeable internal battery
for desk top and mobile version:
Mains adapter:
Input:
Model 1:
Model 2:
Model 3:
Rechargeable Battery:
Classification:
ri-medic table top:
Dimensions:
Weight:
Operating conditions:
Storage conditions:
Airpressure:
IP index:
Calibration:
Safety systems: Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressu-
re to less than 330 mmHg. Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a
blood pressure cycle to less than 180 seconds.
20
Model 1: Table version
Model 2: Mobile stand version
Model 3: Wall version
Adult and pediatric patients (age 3 and above).
Oscillometric
Systolic: 60-270 mmHg
Diastolic: 30-170 mmHg
Heart Rate: 30-200 bpm
MAP: 40-203 mmHg
Typically: 30-40 seconds
BHS (A/A), ESH, AAMI SP10:2002
+- 2% or +- 3bpm, whichever is greater
3 mmHg/sec (Meets AHA Recommendations)
Tube length 70cm/27.56 in (cuff only), total tube length
310cm/122.05 in
200 measurements within a 12-hour
period, 8–12 hours per charge.
Country-specific interchangeable socket plugs/cable available.
External power supply 100-240V 50-60Hz 0,6 A
9V DC 2A
Internal battery 6V DC 5AH
External power supply 100-240V 50-60Hz 0,5 A
9V DC 1,7A
Internal battery 6V DC 5AH
External power supply 100-240V 50-60Hz 0,6 A
9V DC 2A
PC5-6F
6V DC 5AH
Application part type BF,
Defibrillation protected
W 18cm x L 23cm x H 25cm
2200 g
10°C (50°F) to 40°C (104°F) Less than 90% RH
-20°C (-4°F) to 55°C (131°F) Less than 90% RH
800 hPa - 1100 hPa
IPX0
Every two years