Interacoustics Lyra Instructions D'utilisation page 56

Table des Matières

Publicité

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
(a)
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
theLyra is used exceeds the applicable RF compliance level above, theLyra should be observed to verify
normal operation, If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating theLyra.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
(b)
La conformité avec les exigences EMC telles que stipulées dans la norme CEI 60601-1-2 est garantie
si les types de câbles et la longueur des câbles sont comme suit :
Description
Câble EOA
Câble USB
D-0119934-D – 2019/11
Lyra - Instructions d'utilisation - FR
Longueur
2,0 m
2,0m
Blindé ?
Blindé
Blindé
Page 52