Guidance and Manufacturer's Declaration - Electromagnetic Immunity
TheLyra is intended for use in the electromagnetic environment specified below. The customer or the user
of theLyra should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
D-0119934-D – 2019/11
Lyra - Instructions d'utilisation - FR
IEC 60601 Test
level
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
3 A/m
Compliance
+6 kV contact
+8 kV air
Not applicable
+1 kV for input/output
lines
Not applicable
Not applicable
3 A/m
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of theLyra requires
continued operation during
power mains interruptions, it is
recommended that theLyra be
powered from an
uninterruptable power supply
or its battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
environment.