The screen can be mounted on each pump.
The screen can be removed as shown in the fol-
lowing figure.
The following figure shows the overall flange di-
mensions with the protection screen fitted on
pump with ISO flange and pump with CFF flange.
The following figure shows the ISO 250 pump
flange section with the protection screen fitted on
it. As you can see, the overall dimensions do not
change as the inlet screen remains inside the
pump profile.

HEATER BAND INSTALLATION

The heater band model 969-9317 and 969-9327
can be used to heat the pump casing when a
bakeout is needed.
The heater band is available only for the pump
with ConFlat inlet flange (model 969-8932). It is
applied to the upper part of the pump casing, as
shown in the figure, and heats it to a temperature
of about 80° C.
The heater band must be mounted such that there
is perfect thermal contact with the pump wall to ob-
tain fast and efficient heating.
82
TECHNICAL INFORMATION
87-900-945-01(G)