EN
Instructions for use
For Acumed Universal Tray With Dual-Trak Platter:
Storage Case Part
Numbers:
Legacy Instrument Trays:
1
Condition
:
Exposure Temperature:
Exposure Time:
Dry Time:
For Dual-Trak Platter & Lid Only:
Storage Case Part
Numbers:
Condition
1
:
Exposure Temperature and
Time:
Exposure Temperature and
2
Time
:
Dry Time:
1
Use sterilization packaging/wrap and other accessories appropriate for the cycle parameters recommended in these
instructions and in accordance with national regulations. Refer to PKGI-76 at
Aesculap® rigid sterilization containers.
* Sterilization was validated using a STERIS Amsco 3023 Vacamatic Prevacuum sterilizer and KimGuard
KC600 One-Step wrap.
POST-STERILIZATION INSPECTION
• Do not store or use sterile devices if they are not dry.
- Moisture supports the survival of microorganisms.
- Moisture remaining on wrapped or contained products after sterilization could compromise the
sterile barrier.
- Moisture can corrode metal and dull sharp edges.
• Inspect the sterile barrier for signs of damage. Do not use the product if the sterile barrier has been
compromised.
Prevacuum Steam Sterilizer Parameters
Base: 80-0342
Case Lid: 80-0343
80-0344, 80-0345, 80-0785
Wrapped
270°F (132°C)
25 minutes
70 minutes
Prevacuum Steam Sterilizer Parameters
80-0785 or 80-3791 with 80-0786
Wrapped
270°F (132°C)
4 minutes
273°F (134°C)
3 minutes
30 minutes
English (non-U.S.) – EN / PAGE 51