Recommended Neptune System Operating and Storage Temperatures
Table 4: Operating and storage temperature ranges
Condition
Operating temperature
Storage temperature
The Neptune system meets the temperature safety requirements of IEC 60601-
1:2005 in ambient temperatures up to 36.8°C (98°F).
Approved Power Sources
Advanced Bionics has verifi ed Neptune function with the power sources listed
below; behavior when using power sources other than those listed here cannot be
guaranteed.
Table 5: Approved Neptune power sources by geographic region
Power Source
NORTH AMERICA
AAA Battery
Programming Interface
(used during fi tting, only)
EU
AAA Battery
Programming Interface
(used during fi tting, only)
Table 6: Approved Neptune power supplies
Power Supply
AAA Battery
Programming Interface
36 User Guide for the Neptune™ Sound Processor
Minimum
0°C (32°F)
-20°C (-4°F)
Manufacturer
Energizer ®
Energizer
Duracell ®
Advanced Bionics
Energizer
Energizer
Duracell
Advanced Bionics
Rated Voltage
1.0-5.0 V
3.1-5.0 V
Maximum
45°C (115°F)
55°C (131°F)
Model Number
E92 (Alkaline)
L92 (Lithium)
DC2400 (NiMH)
AB-6500
Ultra Plus LR03 (Alkaline)
L92-FR03 (Lithium)
HR03 (NiMH)
AB-6500
Type
Rated Current
DC
250mA
DC
250mA