14. Technical Specifications
Type:
Adjusted mode Non Contact Thermometer NC 200
Mode type:
Adjust mode
Measuring
site:
Forehead
Reference
body site:
Sublingual
Measurement
Body mode: 34.0 - 43.0 °C / 93.2 - 109.4 °F
range:
Object mode: 0.1 - 99.9 °C / 32.2 - 211.8 °F
Resolution:
0.1 °C / °F
Measurement
Body mode:
±0.2 °C, 35.0 42.0 °C / ±0.4 °F, 95.0 107.6 °F
accuracy
±0.3 °C, 34.0 34.9 °C and 42.1 43.0 °C /
(Laboratory):
±0.5 °F, 93.2 94.8 °F and 107.8 109.4 °F
Object mode:
±1.0 °C, 0.1 ~ 99.9 °C / ±2 °F, 32.2 ~ 211.8 °F
Clinical
Repeatability: 0.26 °C
results:
Bias: 0.08 °C
Limits of agreement: 1.13 °C
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
The unit is turned ON and ready for the measure-
ment: 1 short beep.
Complete the measurement: 1 long beep (1 sec.) if
the reading is less than 37.5 °C / 99.5 °F, 10 short
«beep» sounds, if the reading is equal to or greater
than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi» sounds.
Memory:
30 readings recall in the memory mode with a
record of both time and date.
Backlight:
The display light will be GREEN for 1 second,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a reading
less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when
a measurement is completed with a reading equal
to or higher than 37.5 °C / 99.5 °F.
Operating
Body mode: 15 - 40.0 °C / 59 - 104.0 °F
Object mode: 5 - 40.0 °C / 41 - 104.0 °F
conditions:
15 - 95 % relative maximum humidity
NC 200
Storage
-25 - +55 °C / -13 - +131 °F
conditions:
15 - 95 % relative maximum humidity
Automatic
Approx. 1 minute after last measurement has been
Switch-off:
taken.
Battery:
2 x 1.5 V alkaline batteries; size AAA
Battery
approx. 2000 measurements
lifetime:
(using new batteries)
Dimensions: 156.7 x 43 x 47 mm
Weight:
91.5 g (with batteries), 68.5 g (w/o batteries)
IP Class:
IP22
Reference to
ISO 80601-2-56; IEC 60601-1; IEC 60601-1-2
standards:
(EMC); IEC 60601-1-11
Expected
service life: 5 years or 12000 measurements
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.
15. www.microlife.com
Detailed user information about our thermometers and blood pressure
monitors as well as services can be found at www.microlife.com.
EN