Reading 1 - the last reading AQ: Press and release the M-
button 5 to recall the last reading. Number «1» and a flashing
«M» are displayed.
Pressing and releasing the M-button 5 after the last 30 readings
have been recalled will resume the above sequence from reading 1.
10. Error Messages
Measured temperature too high AR: Displays «H» when
measured temperature is higher than 43.0 °C / 109.4 °F in body
mode or 99.9 °C / 211.8 °F in object mode.
Measured temperature too low AS: Displays «L» when
measured temperature is lower than 34.0 °C / 93.2 °F in body
mode or 0.1 °C / 32.2 °F in object mode.
Ambient temperature too high BT: Displays «AH» when
ambient temperature is higher than 40.0 °C / 104.0 °F.
Ambient temperature too low BK: Displays «AL» when
ambient temperature is lower than 15.0 °C / 59.0 °F in body
mode or lower than 5.0 °C / 41.0 °F in object mode.
Error function display BL:
- «Er 0» / «Er 6»: The system has a malfunction.
- «Er 2»: Device is directly placed on the forehead / object.
Keep the measuring distance of 1-5 cm. Do not touch the
bottom side (sensing area) of the measuring sensor.
Blank display BM: Check if the batteries have been inserted
correctly. Also check polarity (<+> and <->) of the batteries.
Flat battery indicator BN: If only «battery» icon is shown on
the display, the batteries should be replaced immediately.
11. Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with alcohol (70%
Isopropyl) to clean the thermometer casing and the measuring
sensor. Ensure that no liquid enters the interior of the device. Never
use abrasive cleaning agents, thinners or benzene for cleaning and
never immerse the device in water or other cleaning liquids. Take
care not to scratch the surface of the sensor lens and the display.
12. Battery Replacement
This device is supplied with 2 new, long-life 1.5V, size AAA
batteries. Batteries need replacing when this icon «battery» BN is
the only symbol shown on the display.
Remove the battery cover BQ by sliding it in the direction shown.
6
Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
13. Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
Transport costs and risks of transport.
Damage caused by incorrect application or non-compliance
with the instructions for use.
Damage caused by leaking batteries.
Damage caused by accident or misuse.
Packaging/storage material and instructions for use.
Regular checks and maintenance (calibration).
Accessories and wearing parts: Batteries, Probe cover
(optional).
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.