Télécharger Imprimer la page

Sanitas SBM 45 Mode D'emploi page 25

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
press the MEM button, the most recent individual meas-
ured values with date and time are displayed in turn.
• You can delete the memory by pressing and holding the
MEM button for 3 seconds. All the values in the current
user memory are deleted after three beeps are output.
• To switch off the device, press the MEM button again or
the START/STOP button or push the touch screen acti-
vation switch to OFF.
• If you forget to switch off the device, it will switch off au-
tomatically after 2 minutes.
8. Error messages/trouble-shooting
In case of faults, the
_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display)
• the cuff is fastened too tightly or loosely (
pears on the display)
• the pump pressure is higher than 300 mmHg (
on the display)
• pumping up takes longer than 160 seconds (
on the display)
• there is a system or device error (
appears on the display)
• the batteries are almost empty
or
ap-
appears
appears
,
,
or
.
25
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
9. Cleaning and storing the instrument
• Clean your blood pressure monitor carefully using a
slightly damp cloth only.
• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
• When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff
tube should not have any sharp kinks.
10. specifications
Model no.
SBM 45
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the up-
per arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown

Publicité

loading