3. Set measuring slide to a value, e.g.
150 cm.
153
152
151
150
130
149
129
148
128
4. Using another measuring rod, measure
the difference from the platform (models
704, 709, 710, 709f, 710f, 799:
) to the measuring slide and note the
floor
difference.
Model 220
5. Undo the screw on the measuring rod
somewhat, move the slide and measur-
ing rod up or down according to the dif-
ference and tighten it back up.
6. Check height again and correct if
necessary.
The measuring rod is also easily readjusted
in this manner if the scale is moved and re-
aligned using the adjustable feet.
from the
Note:
For fine-adjustment of the measuring rod, it
is essential to measure from the floor/plat-
form to the measuring slide. Do not
measure from the floor/platform to the rele-
vant height on the scale. The scale is dis-
placed
a
compensate for the difference in height
between the read-off mark and the
measuring slide.
7. Once
you
screwed the spacer for
the measuring rod to
the scale column and
aligned it, affix the
"Calibration Seal" label
supplied to the spacer
as shown in the diagram above.
few
centimetres
to
have
Operation
How to measure properly...
1. Engage measuring slide in the horizontal
position for measuring.
2. Push measuring slide upwards accord-
ing to the height of the person to be
measured.
For measuring from 130.5 cm to
200 cm (from 51.3 inch to 78.7 inch):
• read-off
mark (1)
1
7
GB