Section 15 EMC Compliance Information for the
LEEP PRECISION Integrated System
Medical electrical equipment needs special precautions regarding EMC and must to be installed and put into
service according to the EMC information provided in the accompanying documents .
Portable and mobile RF communications equipment can affect medical electrical equipment .
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The CooperSurgical LEEP PRECISION Integrated System is intended for use in the electromagnetic environment
specified below. The customer or the end user of the CooperSurgical LEEP PRECISION Integrated System should
assure that it is used in such an environment .
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Compliance
Electromagnetic Environment Guidance
CooperSurgical LEEP PRECISION Integrated
Systems use RF energy only for their internal
Group 1
function . Therefore, RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment .
Class A
CooperSurgical LEEP PRECISION Integrated
Systems are suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
Class A
domestic purposes .
Complies