2.7.2.1 Patient Return Electrode Use
There are two varieties of Patient Return Electrodes that can be used with the
LEEP PRECISION Integrated System:
1
Patient Return Electrode
With Connector
The Cable is built in
(
)
1
: If the Patient Return Electrode with the built-in cable is purchased, the Patient Return
Electrode Cable is plugged directly into socket
(
)
2
: If the Patient Return Electrode and a separate Reusable Cable are purchased, the
Patient Return Electrode is connected to the Reusable Cable, which is then plugged
into socket
.
8
Section 3 LEEP PRECISION Generator Features
•
Electronic power level control
•
Digitally derived waveforms
•
Low inherent leakage rate
•
Patient return monitor disables power automatically if Patient Return Electrode is loose
•
Regulated system performance traceable to NIST Standards
•
Automatic smoke evacuator signal function sends signal to detector for automatic smoke evacuator actuation
when generator is activated
•
Self-monitors all detectors
Classification
Model
LEEP PRECISION Generator
•
Do not get fluid into the LEEP PRECISION Generator. Should any liquid or solid object fall into the unit, unplug
the unit and call Technical Support .
•
The LEEP PRECISION Generator is suitable for intermittent operation with a two-minute ON and six-minute
OFF duty cycle .
•
The LEEP PRECISION Generator is classified as normal equipment (IPX0) according to protection against
ingress of water .
2
Patient Return Electrode
Without Connector
Reusable Cable is purchased
separately from CooperSurgical
.
8
Safety Class
I
11
Type
BF