Section 13 Specifications
13.1 LEEP PRECISION Integrated System
PHYSICAL
Size (W x D x H): . . . . . . 22 .44 inches x 22 .07 inches x 41 .22 inches (57 .0 cm x 56 .06 cm x 104 .70 cm)
Weight: . . . . . . . . . . . . . approx . 88 pounds (40 kg)
ELECTRICAL
Input Voltage
VA Ratings
Fuses
HIGH FREQUENCY OUTPUT*:
Rated output power is 100 watts, 495 kHz ± 5% into 500 ohms, monopolar only .
Minimum & Maximum
Mode
Power Settings
CUT
10 to 100 watts
BLEND
10 to 100 watts
COAG
10 to 100 watts
*Accuracy of output control setting:
For output powers in excess of 10 Watts, the actual power as a function of the load resistance and output
control setting shall not deviate from that shown in the diagrams shown later by more than ±20 percent .
RF Leakage: 100 MA MAXIMUM AT MAXIMUM POWER SETTINGS
13.2 LEEP PRECISION Generator
PHYSICAL
Size (W x D x H): . . . . . . . . . 9 .38 inches x 11 .63 inches x 4 .25 inches (23 .8 cm x 29 .5 cm x 10 .8 cm)
Weight: . . . . . . . . . . . . . . . . 10 pounds, 9 ounces (5 .25 kg)
ELECTRICAL
HIGH FREQUENCY OUTPUT*:
Rated output power is 100 watts, 495 kHz ± 5% into 500 ohms, monopolar only .
Minimum & Maximum
Mode
Power Settings
CUT
10 to 100 watts
BLEND
10 to 100 watts
COAG
10 to 100 watts
*Accuracy of output control setting:
For output powers in excess of 10 watts, the actual power as a function of the load resistance
and output control setting shall not deviate from that shown in the diagrams shown after the
specifications section by more than ±20 percent .
RF Leakage: 100 MA MAXIMUM AT MAXIMUM POWER SETTINGS
120 VAC
1300 VA
10A, 250V, T-Type, Slow Blow,
(Littelfuse® 0218010 or equivalent)
Crest
Factor
1 .2
2 .5
6 .5
Crest
Factor
1 .2
2 .5
6 .5
( Cart with integration unit + Generator + Smoke
Evacuator)
230 VAC
1300 VA
6 .3A, 250V, T-Type, Slow Blow,
(Littelfuse
®
Maximum Open
Circuit Voltage
600
600
4000
Maximum Open
Circuit Voltage
600
600
4000
29