6.Push the retaining ring (in the 11 o'clock
position) and elbow onto the mask frame.
Turn the retaining ring to the 12 o'clock
position.
7.Push the cap onto the pressure measurement
ports.
8.Attach the mask cushion to the mask frame.
9.Fasten the headgear to the mask with the
clips.
10.Check that everything is working properly.
See "Functional check" on page 17.
4 Cleaning and disinfecting
instructions
4.1 Intervals
Daily
Clean the mask in warm water to remove all
residues. To do this the mask can be
disassembled as described in section 3.9
"Disassembling/assembling the mask". Use a
mild cleaning agent (e.g. dishwasher
detergent). After cleaning, rinse all parts
carefully with clean water.
Caution!
For patients with a weakened immune
system or a special medical history,
daily disinfecting or sterilization of the
mask components may be necessary. In
this case, please consult your doctor.
Weekly
Disassemble the mask as described in section
3.9 "Disassembling/assembling the mask"and
clean and sterilize the parts of the mask. See
"Cleaning, disinfection, sterilization" on
page 16.
16
4.2 Cleaning, disinfection,
sterilization
Permissible procedures
Procedure
Parts
Mask cushion
Forehead cushion
Port cap
Mask frame
Coarse adjustment
component
Fine adjustment
component
Elbow
Anti-asphyxia valve
Rotating sleeve
Retaining ring
Headgear clip
Headgear
Quick release
system
Caution!
•Always clean the components of the mask
thoroughly before disinfecting or sterilizing
them as indicated in the above table
•The anti-asphyxia valve is a very sensitive
component. Always handle it with care to
avoid damage or deformation.
Description of procedures
Rinse cycle at 65°C. Then carefully rinse
Dishwasher
all parts with clean water.
Washing machine
Rinse cycle at 30°C.
Clean mechanically, e.g. with a brush,
in dilute solution. After disinfection rinse
Disinfection*
all parts thoroughly with distilled water
and dry thoroughly.
Sterilization
Steam sterilization in devices to comply
(alternative to
with EN 285. Temperature 134 °C,
disinfection)
minimum hold time 3 minutes.
*We recommend the disinfectant
GIGASEPT FF. It has been proven that this
disinfectant is compatible. Follow the
manufacturer's instructions without fail.