TECHNICAL DATA
Desription
Manufacturer
Model
Classification 93 / 42 EEC
Mode
Service Pressure to the turbine
connection : Water
Service Pressure to the turbine
connection : Air
Weight
Operating conditions
Validated life time of the handpiece
Storage conditions
Transport conditions
70
EMS SA, CH-1260 Nyon, Switzerland
AIR-FLOW handy 3.0
Class IIa
Continuous operation
1 to 2.2 bar (1000-2200 hPa)
with a service flow of 50-80 ml/min.
Static pressure 2.7 to 3.5 bar
(2700-3500 hPa)
Approx. 0.150 kg
+10°C to +40°C
30% to 75% relative humidity
700 hPa to 1060 hPa air pressure
Altitude max. 3000 m
800 sterilization cycles
+5°C to +40°C
5% to 85% relative humidity
500 hPa to 1060 hPa air pressure
-10°C to +40°C (ventilated) or
60°C (non ventilated)
75% relative humidity
min. 700 hPa air pressure
CARACTÉRISTIQUES TECHNIQUES
Description
Fabricant
Modèle
Classification 93 / 42 CEE
Mode de fonctionnement
Pression de service au raccord de
turbine: Eau
Pression de service au raccord de
turbine: Air
Poids
Conditions de fonctionnement
Durée de vie validée de la pièce
à main
Conditions de stockage et de
transport
Conditions de transport
EMS SA, CH-1260 Nyon, Suisse
AIR-FLOW handy 3.0
Classe IIa
Service continu
1 à 2.2 bar (1000-2200 hPa)
avec un débit de 50-80 ml/min
Pression statique de 2.7 à 3.5 bar
(2700-3500 hPa).
Approx. 0.150 kg
+10°C à +40°C
30% a 75% d'humidité relative
pression atm. 700 hPa à 1060 hPa
Altitude max. 3000 m
800 cycles de stérilisation
+5°C à +40°C
10% à 95% d'humidité relative
pression atm. 500 hPa à 1060 hPa
-10°C à +40°C (ventilé) ou
60°C ( non ventilé)
75% d'humidité relative
pression atm. min. 700 hPa