6.4 EMC protection
Electrical and magnetic fields often induce interference voltages in the mea
suring circuit. Therefore:
• Use shielded, low‐capacitance measurement cables only (HBM cables
fulfill both conditions).
• Do not route the measurement cables parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with steel
conduit, for example.
• Avoid stray fields from transformers, motors and contact switches.
• Do not ground the transducer, amplifier and indicator more than once.
• Connect all devices in the measurement chain to the same grounded
conductor.
HBM
SLB-700A/06, SLB-700A/12
A2112-3.0 en/de/fr