16
eliminated by a calibration, however, the temperature‐dependent part remains.
The TK
value given in the specifications for the transducer therefore does not
c
apply for the cable and transducer combination when connection is with four‐
wire configuration, the cable percentage must be added.
6.3 Shortening the cable
As the transducer connection uses a six‐wire configuration, you can shorten
the 6‐wire transducer cable without affecting the measurement accuracy.
6.4 Cable extension
The cable of a six‐wire force transducer can be extended with a cable of the
same type.
Only use shielded, low‐capacitance measurement cables for extension. En
sure that connection is perfect, with a low contact resistance.
6.5 EMC protection
Electrical and magnetic fields often induce interference voltages in the mea
suring circuit. Therefore:
• Use shielded, low‐capacitance measurement cables only (HBM cables
fulfill both conditions)
• Do not route the measurement cables parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with e. g. steel
conduit.
• Avoid stray fields from transformers, motors and contact switches.
• Do not ground the transducer, amplifier and indicator more than once.
• Connect all devices in the measurement chain to the same grounded
conductor.
HBM
S2
A3181-1.0-3.0 en/de/fr