Select the display unit "Uni" before you define the values for the
parameters SPx, nSPx, FSPx, ASP, AEP. This avoids rounding errors dur-
ing internal conversion to other units and enables exact setting of the
values. When delivered: mm.
If no button is pressed for 15 s during the setting procedure, the unit
returns to the Run mode with unchanged values.
The unit can be electronically locked to prevent unwanted adjustment
of the set parameters: Press the two programming buttons in the Run
mode until
is displayed.
On delivery: Unlocked (
When the unit is locked,
change parameter values.
Set-up / Operation
After mounting, wiring and programming check whether the unit
operates correctly.
NOTE: If output 2 is connected to an analogue input card, the para-
meters of the output must be set first (OU2 = I or OU2 = U).
Otherwise the input card can be destroyed.
Align the unit (laser beam) to the object to be detected. If correctly set
up, the distance to the object is indicated.
Faults displayed during operation:
++
too much light, e.g. reflective surfaces
--
not enough light
far
measured object outside the measuring range > 10 m
near
measured object outside the measuring range < 0.2 m
SC1
short circuit in switching output 1
SC2
short circuit in switching output 2
SC
short circuit in both switching outputs
Life of the laser diode: 50 000 hours.
26
is displayed. For unlocking press the buttons until
).
is briefly displayed when you try to